Blog · By IVF Automation team · 16 Oct 2026
Daily IVF lab QC logs: what to track
If an inspector asked for last month's incubator readings today, could you produce them in five minutes? Daily quality control logs are the backbone of a well-run IVF lab. They show that equipment and environment were checked, problems were noticed early and someone acted on them. They are also among the first documents an inspector or auditor asks to see. This article covers what to log, how to structure the log and how to review it. It is operational only. The acceptable ranges and actions for your lab must come from your lab director, your SOPs and each equipment manufacturer's instructions, not from this article.
What the guidance says
ESHRE's Revised guidelines for good practice in IVF laboratories (2015) set the documentation expectations behind a QC log:
- Accepted ranges for all measured parameters should be determined and recorded.
- If a measurement is out of range, it should be corrected and the correction verified.
- Equipment validation, calibration, maintenance and repair must be documented and the records kept.
- Malfunctioning equipment should be labelled "out of use".
- Critical equipment, including incubators and cryostorage units, should be continuously monitored with alarm systems.
- Critical equipment must have automatic emergency backup power.
ESHRE published an updated set of Recommendations on Good Practice in the IVF laboratory in Human Reproduction in 2026. Check the current version. In India, the ART (Regulation) Act 2021 (section 23) also requires clinics and banks to keep detailed records and make them available for inspection.
Principles of a good QC log
- Every entry has a date, time and initials. An entry without a name cannot be trusted later.
- Record the actual reading, not just a tick. "37.0" tells a reviewer far more than "OK".
- Show the acceptable range on the form. Staff can then see instantly when a reading is out of range.
- Every out-of-range reading has a recorded action. What was done, by whom and when, and the reading after the action.
- Logs are reviewed and signed off. Daily by the person in charge, and periodically by the lab director.
- Corrections are visible. Strike through once and initial, or use a system with an audit trail. Never overwrite.
Our free daily lab QC log follows these principles and can be printed or used digitally.
1. Incubators
For each incubator, identified by a unique ID, log at the times your SOP specifies:
- Displayed temperature and, if your SOP uses one, the reading from an independent calibrated thermometer or probe.
- Displayed gas concentrations the incubator controls, and independent verification readings when your SOP requires them.
- Humidity status where relevant to the model.
- Door alarms or error messages since the last check.
- Any cleaning or maintenance done that day.
Recording both the displayed and the independent reading helps reveal sensor drift early.
2. Temperature-controlled equipment
List every temperature-controlled item: heated stages and plates, warming blocks, refrigerators and freezers for media and consumables, and room temperature in the lab itself. For each, log the reading, the acceptable range from your SOP and the initials of the person checking.
3. Gas supply
Gas interruptions can affect several items at once. Log:
- Cylinder or manifold pressures for each gas, and which cylinder is in use.
- Cylinder changes, with date, time, cylinder identifier and who changed it.
- Remaining stock of full cylinders, so that reordering happens in time.
- Any automatic changeover events or alarms.
4. Cryo storage
For storage dewars and tanks:
- Liquid nitrogen level or weight, measured in the way your SOP specifies.
- Refill dates and quantities.
- Status of level or temperature alarms.
- Physical checks such as frost or damage on the vessel.
- Liquid nitrogen supply stock and delivery dates.
Inventory movements belong in your storage register rather than the daily log, but cross-reference them. Our cryo renewals tool helps with renewal tracking.
5. Alarms and monitoring systems
Alarms only protect you if they work. Log:
- A daily check that the monitoring system is online and receiving data from each device.
- Scheduled alarm tests, with the result and who received the test notification.
- Every real alarm: time, device, cause, action taken and time resolved.
- The current on-call contact list, checked whenever the roster changes.
6. Environment and facility
Depending on your lab design and SOPs, consider logging:
- Room temperature and humidity.
- Air handling or filtration status and filter change dates.
- Pressure differential readings if your lab uses them.
- Power events: outages, UPS and generator switchovers and their duration.
7. Cleaning and housekeeping
Log each scheduled cleaning task, the agent used, who did it and when. Link the cleaning schedule to your SOPs so that deep cleans and incubator cleans are never missed.
8. Consumables and media
Operational records here usually include lot numbers and expiry dates of items in use, the date each item was opened, and storage conditions. Your SOP defines which items need to be tracked and how.
9. Equipment service and calibration
Keep a separate equipment file but note on the daily log when a service or calibration visit happens. Record the vendor, engineer, work done and when the next visit is due. Our guide to choosing equipment vendors explains what to ask for in service contracts.
How to structure the log
Two common approaches work well:
- One sheet per day covering every item. Good for small labs and fast review.
- One sheet per device per month, with a row for each day. Good for spotting trends on a single device.
Whichever you choose, keep the same layout every day so that missing entries are obvious at a glance.
Paper or digital?
Paper is simple and always available, but hard to search and trend. Digital logs with an audit trail make trend review and inspection retrieval much easier. Many labs combine both: digital monitoring for continuous data and a paper or tablet checklist for manual checks. Whatever you use, make sure records are backed up and retained for the period your legal and accreditation requirements specify.
Reviewing the logs
- Daily: the person in charge checks the log is complete and every out-of-range reading has an action.
- Weekly: look for trends such as slow drift on one incubator or frequent gas changes.
- Monthly: the lab director reviews and signs off, and recurring issues are raised with vendors or management.
Turn repeated issues into measurable indicators. The Vienna consensus (2017), from ESHRE and Alpha Scientists in Reproductive Medicine, defines 19 laboratory indicators, 12 of them key performance indicators, and recommends that each lab choose and track its own set. Our KPI page has examples of operational metrics.
Sources
- ESHRE Revised guidelines for good practice in IVF laboratories (2015)
- ESHRE Recommendations on Good Practice in the IVF laboratory, Human Reproduction (2026)
- The Vienna consensus, Reprod Biomed Online 2017
- ART (Regulation) Act 2021, section 23
Quick checklist
- Every device has a unique ID
- Acceptable ranges printed on the form, from your SOPs
- Actual readings recorded with time and initials
- Out-of-range actions documented
- Gas, cryo and alarm checks included
- Daily review and monthly sign-off
- Logs backed up and retained
Your next step
Open our free daily lab QC log and start using it on tomorrow's morning round. To get practical lab-ops guides like this every week, subscribe to IVF Automation Weekly using the form below.
Good logs make inspections straightforward; see our ART Act compliance checklist. For staffing, use the jobs board; for equipment and service providers, the vendor directory; and for training opportunities, the events calendar. This is operational information only; see our disclaimer.
General operational information only; not medical, clinical or legal advice. See our disclaimer.